Loading... Please wait...

Categories

Our Newsletter


AnalytikJena CL-3000 UV Crosslinker

  • Image 1
  • Image 2
Price:
$2,321.97
Weight:
22.00 LBS
Availability:
Built to order, 3-week lead time.
Shipping:
Calculated at checkout


Product Description

AnalytikJena's Model CL-3000 UV Cross-linker supplies uniform intensity overhead UV exposure of materials. It utilizes shortwave ultraviolet energy to bond DNA to a medium. Ultraviolet Crosslinking for attaching nucleic acids to a membrane takes seconds as compared to oven baking.  AnalytikJena Crosslinkers offer researchers an instrument to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. A unique UV sensor continually measures the UV energy and automatically adjusts to variations in UV intensity that occurs as the UV tubes age. This same UV sensor feedback measurement system allows you to set UV sample exposure, which automatically deactivates the UV sources when the set UV energy dose has been achieved.

Applications

  • UV induced mutations,
  • Membrane crosslinking,
  • Biomolecular crosslinking,
  • UV dose calibrated to a NIST traceable standard,
  • UV sterilization and sanitization,
  • UV curing.

Features

  • Preset and manual controls for ultraviolet or time exposures,
  • Large LED display with touch panel displays time or energy settings,
  • Window on the door allows viewing of the process yet blocks the UV radiation,
  • Small footprint for benchtop use,
  • Six 8-watt ultraviolet tubes provide uniform overhead ultraviolet illumination.

Model

P/N

Wavelength

115V

230V

CL-3000

849-95-0615-01

849-95-0615-02

254nm

CL-3000M

849-95-0615-03

849-95-0615-04

302nm

CL-3000L

849-95-0615-05

849-95-0615-06

365nm

Specifications

Technical Specifications

CL-3000

CL-3000M

CL-3000L

Wavelengths

254 nm

302 nm

365 nm

Bulbs

6 x 8 Watt

Energy

0000.1 - 9999.9 mJ/cm² (0 - 10 J/cm²)

Time

000:01 - 999:59 (>300 J/cm²)

Temperature

15⁰C - 35⁰C

Humidity

70% Non-Condensing

Altitude

up to 3,000M (9,842 ft.)

Sound Level

≤ 50 dba

Housing Surface Temp

≤ 30⁰C

Startup Time

< 1 sec

External Dim (L x W x H)

41 cm x 40 cm x 26.5 cm

Internal Dim (L x W x H)

35 cm x 27 cm x 16 cm

Weight

6.8 Kg: 15 lb

Operating Power

100 - 115VAC & 230VAC 50/60Hz

Certifications

CE, RoHS (CSA In Process)

Optional Accessories

  • UV meters - handheld UV Meter used to ensure the maximum operating efficiency of UV light sources with the measurement of a UV illumination system or lamp's UV output - the most effective way to monitor the useful life of any UV light source. 

fff.jpg

  •  UV blocking eyewear Our line of safely eyewear will primarily protect against intense visible and UV light from the artificial and natural sources and can increase visibility in bright conditions.

eyewear.jpg

---

Warranty Information

AnalytikJena's products are guaranteed to be free of defects in materials, workmanship and manufacture for one (1) year from date of purchase, except as follows:
- Consumable and disposable parts are guaranteed to be free from defects in manufacture and materials for ninety (90) days from date of purchase.
- UV tubes and filters are guaranteed to be free from defects in manufacture and materials for ninety (90) days from date of purchase.
- AnalytikJena Transilluminators are guaranteed to be free from defects in manufacture and materials for two (2) years from date of purchase.

If equipment failure or malfunction occurs during the warranty period, AnalytikJena shall examine the inoperative equipment and have the option of repairing or replacing any part(s) which, in the judgment of AnalytikJena , were originally defective or became so under conditions of normal usage and service.

No warranty shall apply to this instrument, or part thereof, that has been subject to accident, negligence, alteration, abuse or misuse by the end-user. Moreover, AnalytikJena makes no warranties whatsoever with respect to parts not supplied by AnalytikJena or that have been installed, used and/or serviced other than in strict compliance with instructions appearing in this manual.

In no event shall AnalytikJena be responsible to the end-user for any incidental or consequential damages, whether foreseeable or not, including but not limited to property damage, inability to use equipment, lost business, lost profits, or inconvenience arising out of or connected with the use of instruments produced by AnalytikJena . Nor is AnalytikJena liable or responsible for any personal injuries occurring as a result of the use, installation and/or servicing of equipment.

This warranty does not supersede any statutory rights that may be available in certain countries.

Find Similar Products by Category

Write your own product review

Product Reviews

This product hasn't received any reviews yet. Be the first to review this product!